cAUtIon: do not remove the Flexscope cable from the System Workstation by pulling on the
cable� Improper removal of the Flexscope cable may damage the System Workstation�
dispose of the LithoVue Flexscope in accordance with hospital, administrative and/or local
government policy�
5�5 SAFELy SHUttInG doWn tHE SyStEM WorKStAtIon
Follow these steps to shut down the System Workstation at the end of a procedure or during a procedure:
1. If a restart is required during a procedure, press the "Power" button on the front of the System
Workstation to power it down. Another press of "Power" button will restart the system.
2. To shut down the system at the end of a procedure:
a) Remove the LithoVue™ Flexscope from the patient and prepare for disposal, as described in
the LithoVue Single-Use Digital Flexible Ureteroscope Directions for Use.
notE: the LithoVue Flexscope may be unplugged from the System Workstation before or
after the System Workstation has been powered down.
b) Press the "Power" button on the front of the System Workstation to power it down.
c) Clean the System Workstation after each use as described in Section 6 "Cleaning and
Maintenance".
5�6 dISPoSAL oF tHE ProdUct, AccESSorIES, And PAcKInG MAtErIALS
Dispose of all products, accessories and packaging materials in accordance with hospital, administrative
and/or local government policy.
6 cleaning and Maintenance
6�1 cLEAnInG tHE SyStEM WorKStAtIon
cAUtIon: to avoid the potential for electric shock, disconnect the Power cord from the hospital's
electrical outlet before cleaning the System Workstation�
The System Workstation may be cleaned with mild, non-abrasive cleaning solutions commonly used in
the hospital environment. The System Workstation has been tested and demonstrated to be compatible
with the following cleaning solutions:
• 70% Isopropyl alcohol
• Ammonium Chloride 10%
• Sodium Hypochlorite 6%
Do not spray the Touch PC with cleaning solution. Apply the cleaning solution to a cleaning cloth and wipe
all external surfaces.
cAUtIon: to avoid System Workstation damage, do not allow fluids to enter the touch Pc
enclosure, Power cord connections, connector cable receptacle, or other component/accessory
connections�
cAUtIon: Using cleaners and disinfectant solutions that contain long-life surfactants can leave
conductive residues on the contacts in the Workstation receptacle� the conductive residues
can lead to malfunctions of the System Workstation� to prevent electrical malfunctions due to
conductive residue buildup on the contacts, do not use cleaners and disinfectant solutions that
contain long-life surfactants�
User's Manual
35