7.3.8 Screwing the bottom plate to the connection adapter

27
Complete the following steps to screw the magnetic bottom shell to the connection adapter:
1) Insert the magnetic bottom shell into the medial side rail from the inside.
2) Screw the magnetic bottom shell and the connection adapter to each other from the outside
with the cap screws (M6) (see fig. 27).
3) Tighten the cap screws with a torque of 10 Nm.
4) Re-tighten all screws by turns after 30 minutes and secure with 636K13 adhesive.
INFORMATION: When securing the screws, only loosen one screw at a time, apply the
adhesive to the thread and tighten the screw to the specified tightening torque.
The connection adapter can be installed in three positions depending on the flexion contracture.
Flexion contracture
None
Up to approx. 5 degrees
Greater than 5 degrees
The following conditions have to be met to ensure there is a strong connection between the mag­
netic bottom shell, medial side rail and connection adapter:
All six screws always have to be used.
The screws have to be tightened by turns several times to a torque of 10 Nm until all screws
turn no further.
The screws are re-tightened after 30 minutes and secured with 636K13 adhesive as
described above.

7.4 Finishing the prosthetic socket

Installing the socket adapter
The socket adapter can be selected from the list of possible combinations (see Page 27).
Use the supplied countersunk head screws (M6x12 and M6x16) to install the socket adapter on
the connection adapter.
A medial or lateral installation position can be used for the socket adapter depending on the
residual limb position. When socket adapters with a centring collar are used, two 4Y19 pressure
plates have to be installed between the socket adapter and connection adapter. When a socket
44
Recommended screw position
Low flexion position of the prosthetic socket
Install the connection adapter in the anterior
position.
Moderate flexion position of the prosthetic
socket
Install the connection adapter in the middle.
High flexion position of the prosthetic socket
Install the connection adapter in the posterior
position.