<5> The fifth screen in the "editing splice mode" menu:
Parameter
Arc1 Power
Arc1 Time
Arc2 Power
Arc2 Time
Arc2 ON Time
Arc2 OFF Time
©Copyright Abacanto Digital SA.
Arc discharge can be separated into two stages.
Arc1 Power is the first stage. This sets Arc1 Power.
Sets Arc1 time.
Caution If Arc1Time is set 1 sec. or less and Arc2
Power is set to "OFF", the splice may
break during proof-test stage.
Arc2 is the second arc discharge stage. This sets
Arc2 Power.
Sets the total Arc2 time. Usually set this value to
"OFF". It is possible to set a very long arc time.
However, when the total of the Arc 1 time and Arc2
time exceeds 30 seconds, always adjust the function
[Arc2 ON Time] and [Arc2 OFF Time] to weaken
the arc power. A continuous arc over 30 seconds,
without weakening the arc power, may damage the
arc discharge unit.
During Arc2 discharge, arc power can be pulsed by
turning the arc on and off. This sets the amount of
time that Arc2 is ON.
Sets Arc2 OFF Time during Arc2. When the
Arc2 discharge is intermittent, re-arc discharge is
also intermittent.When continuous re-arc discharge
Manual de usuario AD300