refer to parameter settings of the other splice modes stored in the
database area.
6.3 Heating Operation
Fiber protection sleeve does not shrink completely
z Extend the heating time. Refer to [Referring or Select/editing Heater
Mode] (chapter 3.2.3).
Fiber protection sleeve adhered to heating plate after shrinking
z Use a cotton swab or a similar soft tip object to push and remove the
sleeve.
Method to cancel heating process
z Press
6.4 Senior Settings
Method to lock "selection" or "editing" of Splice or Heater mode
z Refer to [Menu Lock]
6.5 Other Functions
Method to hide messages on [READY] screen
Change the fiber image from X/Y view to X magnified view or Y
magnified view by pressing
Too many repetitions until "Test Finish" indicated in [Discharge
Calibration]
z The splicer needs to repeat fiber preparation and arc calibration after
replacing the electrodes or when the environmental conditions change
drastically. After "Step 1 and Step 2 of discharge calibration
completed is continuously displayed for 3 times, the Fusion Splicer
automatically exits from the discharge calibration interface and enters
"Splice mode" menu.
"Test Finish" is never indicated after many repetitions in [Discharge
Calibration]
z Execute [Stabilizing Electrodes] function in [Maintenance menu]. If
the splicer still does not indicate "Test Finish", replace the electrodes.
(Сhapter 5.5).
No arc power change after [Discharge Calibration]
z An internal factor is calibrated and adjusted for the specific arc power
selected. The displayed arc power in each splice mode does not
change.
z The calibration results affect all splice modes.
z Refer to [splice procedure].