cursor to [sleeve type]; then press
the"FP-03,FP-03 Ny-08,FPS-01-250-20,BLANK"option;
press
double press
"××" (such as Number"10") is deleted, that is, the file is
changed back into blank heating mode (BLANK).
Note3: The user may also modify some parameters of the newly
copied heating file numbered "××" (such as Number"10").
Refer to Chapter 3.2.3 [selecting/editing heating
mode]-<2> editing heating parameter files.
3.2.5 [discharge calibration]
<1> Electrode wears and glass adhesion may result in
changes in arc power. Also, the center position of arc
discharge sometimes shifts to the left or right. In this
case, it is necessary to perform an arc power calibration
to eliminate these problems. [discharge calibration] of
this machine has the function to test the current arc
intensity and revises it into the standard intensity.
<2> Discharge calibration is divided into two steps: "Step I
of Calibration", "Step II of discharge calibration".
In the "Splice Mode menu", press
©Copyright Abacanto Digital SA.
to make the cursor select "BLANK"; then
, and the copied heating mode numbered
the cursor to [discharge calibration]; after the fiber is
prepared and arranged, press
interface of "Step I of discharge calibration"; and is
automatically started.
Discharge calibration starts from "Step I of discharge
calibration". See the following picture.
Manual de usuario AD300
AD Instruments
to enter
to move
to enter the