1
Feature
1
Bed mount
2
Purge cassette door release
3
VGA OUT
4
USB connector
5
Service
6
AC fuses
7
AC plug
8
Power switch
9
Equipotential ground stud
10 Ethernet jack
Automated Impella Controller
2
Description
Metal bracket on the back of the controller; attaches controller to the cart or bed
Button located on the left side of the controller; press to open the purge cassette door
Connection for connecting the controller to another monitor to slave the display
Interface for data transfer by Abiomed maintenance or service personnel
Connection used by Abiomed maintenance or service personnel
Electrical safety device in the event of current overload
Connection point on the controller for the AC power cord
Button that turns the controller on or off
• ON: Press and hold the power switch for 2 seconds
• OFF: (1) Disconnect the Impella Catheter from the Automated Impella Controller
(2) Press and hold the power switch for 3 seconds
(3) A pop-up confirmation box will appear
(4) Press OK using the selector knob to confirm that the controller should be turned off
NOTE: Do not hold down power switch for more than 30 seconds during operation--this initiates an emergency shutdown
Used to ground the Automated Impella Controller according to hospital procedures
Connection for downloading data
3
10
9
8
7
4
5
6
5