Push the Save button [1] to store the setting from
PROGRAM 1 in the memory.
•
The list displays the names of the created programs.
•
There are 50 positions available
NOTE
•
It's possible to overwrite stored programs.
•
Go to Settings to clear the complete memory. Only
the created programs will be deleted.
5.7.5
Load the created programs
•
Select with push button [1] Select Program in the
main menu.
•
Select "User Programs" with the selection button [1]
•
Select a program in the list with the central controller
[11]
•
Confirm the selection with the confirmation button [10]
•
The program will be loaded on both channels
5.8
Laser test
•
Turn the unit on and adjust the parameters
•
Press button [5] to start the laser emission
•
Hold the laser probe above the laser tester [7]
•
The information pop-up screen informs you that emission is present.
•
After a short period the information disappears.
5.9
Therapy application
Once you have pressed START, the TIMER will set off and this can be seen by the timer symbol and the
arrow () in correspondence with the channel (a square ◼ indicates that the channel is not active).
For example, we can see that channel 1 is not active while channel 2 is performing a program.
The frame around Channel 2 indicates that the visualized program is related to channel 2.
Press the corresponding function key in order to see the program which has been loaded in the other
channel.
Countdown and emission continue until:
•
The set time has elapsed: the system will then emit an acoustic signal.
•
The STOP button has been pressed once: the cycle being emitted will be put on Pause , push
STOP again to end the treatment.
The machine emits a BEEP at 3 second intervals to indicate that emission is in progress during therapy
application.
✓
Save
Return
Create Program
Freq.: 6500Hz
Time: 00:20 s
Dosage: 6 J/ cm²
PROGRAM 1
User Programs
Return
Select Program
01..............
02..............
03..............
04..............
05..............
Protocol-05
06.............
07.............
EN
15