Magnetic field at the
network frequency
(50 / 60 Hz)
IEC 61000-4-8
NOTE : UT is the a.c. mains voltage prior to application of the test level.
ENDOLASER 422 is designed to work in the electromagnetic environment specified below. The client or user of
ENDOLASER 422 should ensure that it is used in this environment
Portable and mobile RF communications equipment should not be used closer to any part of, including cables, than
the recommended separation distance calculated with the equation applicable to the transmitter frequency.
Immunity test
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
Immunity to
proximity fields
from wireless RF
communication
devices
IEC 61000-4-3
EN
28
70% U
for 25 cycles
T
0% U
for 250 cycles
T
3 A/m
Guide and declaration of the manufacturer - electromagnetic immunity
Trial level of the IEC 60601
3 Veff
from 150kHz to 80 MHz
3 V/m
from 80 MHz to 2,5 GHz
TETRA 400
380 – 390 MHz
GMRS 460
FRS 460
430 – 170 MHz
LTE Band 13, 17
704 – 787 MHz
GSM 800/900, TETRA 800, iDEN 820,
CDMA 850, LTE Band 5
800 960 MHz
GSM 1800; CDMA 1900; GSM 1900;
DECT; LTE Band 5
1700 – 1990 MHz
Bluetooth, WLAN, 802.11 b/g/n, RIFD
2450, LTE Band 70
2400 – 2570 MHz
WLAN 802.11 a/n
5100 – 5800 MHz
it is recommended to use
70% U
for 25 cycles
ENDOLASER 422 with a UPS or
T
batteries.
0%
U
for
250
T
cycles
Not applicable, the
Network frequency magnetic fields
device
does
not
should have characteristic levels of a
contain components
typical location in a commercial or
susceptible
to
hospital environment.
magnetic fields.
Level of
compliance
3 Veff
3 V/m
27 V/m
27 V/m
28 V/m
28 V/m
9 V/m
9 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
28 V/m
9 V/m
9 V/m
Recommended
separation distance d:
d= 30 cm
d= 30 cm
d= 30 cm