dow in ZEN-Xi.
If the sensor is connected, "READY" appears in green in window MyRay
light on ZEN-X will become green.
At this point, the ZEN-X system is ready to receive an x-ray image.
If the sensor is NOT connected, "SENSOR NOT CONNECTED" appears in red in window iCapture Monitor after a few
seconds and the light on ZEN-X will become yellow.
To quit this condition, connect a sensor to the interface or switch over to position Door closed.
Stand-By
This position switches ZEN-Xi to Stand-by.
The light on ZEN-Xi becomes blue and the system switches to stand-by. When turned to On, the system is immediately
read to operate.
Always use the Stand-By position to disconnect the sensor's connector.
Switching from Stand-by to On is detected by the electronic interface that automatically opens the connection between
the sensor and computer or workstation.
Door closed
As ZEN-Xi is physically attached to the dental chair instrument board, dirt and sprays may hit the connector when the
sensor is not connected to the interface.
Turn the sensor cradle until the opening is covered by the relative part for hygienic reasons
and to avoid damaging electronic parts.
If the door is left opened without the sensor, the light will become yellow and flash and "sensor not connected" will ap-
pear in window iCapture Monitor in red.
Extraction
The sensor cradle can be taken out of the interface in this position so that cleaning and disinfecting operations can be
performed.
ZEN-X / ZEN-Xi
- OPERATING INSTRUCTION
Driver TWAIN
after a few seconds and the
®
®