Instructions for Use
The pump is heavy and must be handled by means of suitable moving and
WARNING
handling tools. During handling it is always recommended the simultaneous
use of four eyebolts.
Do not install or use the pump in an environment exposed to atmospheric agents
(rain, snow, ice), dust, aggressive gases, or in explosive environments or those with
a high fire risk. During operation, to obtain the declared functioning specification,
the ambient temperature must be between 0 °C and +85 °C and the humidity
between 0 and 95% (without condensation).
The pump should be kept sealed with its pinch-off until it is ready for
CAUTION
attachment to the vacuum system.
WARNING
After removing the shipping flange, to avoid injury, never connect the high
voltage to the pump before it is installed into the system and all the inlet
flanges are properly connected or blanked off.
The pump operation is optimized using one of the special Agilent controllers only.
The safety specifications agreement using the pump is guaranteed using the
CAUTION
Agilent controller only.
192
Agilent VacIon Plus 800 User Manual