Inspection Procedure

VacIon Plus pumps are evacuated, baked out, sealed and leak-checked prior to
shipping. The following information and procedures can be used to evaluate the
vacuum integrity of a VacIon Plus pump before installation.

Visual Inspection

Inspect the pump for physical damage which may have occurred during shipment.
Inspect the pinch-off seal (that is covered by a rubber cap). If it is open, the pump is
at atmospheric pressure.
The pinch-off seal is extremely sharp. Be careful when opening. Watch your
WARNING
fingers.
A VacIon Plus pump that has been exposed to atmosphere during shipment, or
while in storage, will operate properly if it has not been damaged.
The pump is not harmed by such exposure, although it is good practice to keep it
under vacuum when not in use, in order to avoid the dust contamination and the
accumulation of water vapor from the environment.