Instructions for Use
VacIon Plus 800 mounting
For installations in vertical in which the pump is not resting at the base or
WARNING
for installation in a position other than vertical, the pump, due to its weight,
must be conveniently supported. It is recommended the always use four
attachment points.
The VacIon Plus 800 pump can be installed in any position, if conveniently
supported. For convenience it is normally mounted in a vertical position with the
inlet flange at the top, using as a support the brackets under the pump. For
installations in vertical in which the pump is not resting at the base or for
installation in a position other than vertical, the pump, due to its weight, cannot be
mounted cantilevered, but must be suitably supported by the supports (bars or
plates of support, whose geometry depends on the type of system on which the
pump is mounted). Such supports can be coupled to the pump via the studs (with
M12 thread) present on both sides of the pump. It is recommended to always use
four lifting points simultaneously. For detailed information about the pump
installation, see the appendix "Technical Information".
Figure 33 VIP 800 showing M12 threaded studs
Agilent VacIon Plus 800 User Manual
197