11. Manipulate the device to allow rinse water to drain from the
lumen.
12. Visually examine each device for cleanliness.
If visible soil remains, repeat cleaning procedure until
13.
the device is thoroughly clean.
Automatic Cleaning
1. Ensure all pre-processing instructions are followed prior to
cleaning.
2. Clean the devices via the Automatic Cleaning parameters below.
PHASE
RECIRCULATION
Pre-Wash
1
Enzyme
Wash
Wash 1
Rinse 1
Pure Rinse
Drying
3. Manipulate the device to allow rinse water to drain from the
lumen.
4. If visible moisture is present dry the instrument with a clean,
lint-free towel.
5. Visually examine each instrument for cleanliness.
6. If visible soil remains, repeat cleaning procedure until the
device is thoroughly clean.
Inspection/Maintenance
Proper care and handling is essential for satisfactory
performance of any surgical device. The steps in these
instructions for use should be taken to ensure long and trouble-
free service from all your surgical devices. Inspect devices
before each use for broken, cracked, tarnished surfaces,
movement of hinges, and chipped or worn parts. If any of these
conditions appear, do not use the device. Return device to
authorized repair service center for repair or replacement.
Before sterilizing, lubricate the device with instrument milk or a
steam permeable/water soluble lubricant, following the lubricant
manufacturer's instructions. Check for smooth action of all
moving parts. Figure 9
Immediately proceed to recommended steam sterilization. Do
not store instrument prior to sterilization.
Let devices drip dry for three (3) minutes before packaging for
sterilization.
MINIMUM
TIME
15 Seconds
1 Minute
2 Minute
15 Seconds
10 Seconds
N/A
WATER
TEMPERATURE
Cold Tap Water
1°C - 16°C
(33°F - 60°F)
Hot Tap Water
43°C - 82°C
(110°F - 179°F)
Tap Water
43°C - 82°C
(110°F - 179°F)
Tap Water
43°C - 82°C
(110°F - 179°F)
Purified Water
43°C - 82°C
(110°F - 179°F)
N/A
3
DETERGENT TYPE
AND CONCENTRATION
(IF APPLICABLE)
N/A
• Detergent: pH-
neutral/enzymatic
detergent
• Concentration: Per
the detergent
manufacturer's
recommendations
• Detergent: pH-neutral
cleanser
• Concentration: Per
the detergent
manufacturer's
recommendations
N/A
N/A
N/A