2.1 Control Panel (see Figure 1, Diagram/Contents)
The Control Panel contains the following LED indicators: Motor On, Standby, Filter Life, and Service.
Read all instructions before operating the Bovie
accessories. Failure to do so may result in damage to the Bovie
and/or personal injury.
POWER OFF / ON
To power up the Bovie
®
grounded electrical outlet and the power receptacle at the bottom of the Bovie
Evacuator. Flip the Power Off / On Button to "ON" to apply power to the unit. Once power has been applied,
the Standby Indicator LED illuminates the color amber. Turn the Bovie
Evacuator main power off by either flipping the Power Off / On Button to "OFF" or by unplugging the Power
Cord from either the power receptacle on the Bovie
grounded electrical outlet.
SUCTION ON / STANDBY BUTTON
Press the Suction On/Standby Button to shift between two (2) modes: On or Standby. The Suction On
LED illuminates the color green indicating active suction and the Standby LED illuminates the color amber
when in standby.
Filter LIFE INDICATOR
The Filter Life Indicator on the Control Panel provides a visual indication of the status of the filter life of the
filter in use.
Install an unused filter into the Bovie
instructions (section 2.3). When the Motor On LED is illuminated, the Filter Life Indicator will light up the
uppermost green LED indicating 100% Filter Life. The indicator will progress through subsequent colors,
from green LEDs to amber LED as time elapses, and finally to red to indicate the filter has expired and
requires replacement. When the maximum filter life is expended: The unit will cease operation in one
(1) hour or when the unit is manually powered off, whichever occurs first. The smoke evacuator will not
resume functionality until a new filter is installed. The filter has
a 6 hour lifetime which is automatically tracked by the Bovie
Smoke Shark Surgical Smoke Evacuator.

2.2 Set-Up and Operation

1.
Mount the Bovie
Smoke Shark Surgical Smoke Evacuator
®
using the wall mounting kit (SE03-M). Secure the wall
mounting kit to the wall by drilling two 3/8" diameter holes
in wall, installing the provided hardware into the holes,
and secure mounting plate to the provided hardware.
The Bovie
Smoke Shark Surgical Smoke Evacuator can
®
be secured to the wall mounting kit (SE03-M) by sliding
the two screws protruding from the rear of the Bovie
Smoke Shark Surgical Smoke Evacuator (Figure 2.) into the
grooves found on the mounting kit (SE03-M).
2.
Attach the Power Cord to the power receptacle on the bottom of the Bovie
Smoke Evacuator and into a grounded electrical outlet. Route Power Cord to prevent a tripping
hazard or crimping of cords, which could cause unreliable operation or electric shock.
Customer/Technical Services: 800.251.3000 (United States) or +1 615.964.5532 (International)
Smoke Shark Surgical Smoke Evacuator, connect the supplied Power Cord to a
Smoke Shark Surgical Smoke Evacuator as per the installation
®
OPERATING INSTRUCTIONS: SECTION 2.0
Smoke Shark Surgical Smoke Evacuator or installing
®
Smoke Shark Surgical Smoke Evacuator
®
®
Smoke Shark Surgical Smoke Evacuator or the
®
®
®
P000027200 Rev A Bovie
Smoke Shark Surgical Smoke
®
Smoke Shark Surgical Smoke
Figure 2
Smoke Shark Surgical
®
Smoke Shark III Operator's Manual
®
Page 10