Guidance and Manufacturer's Declaration - Electromagnetic Immunity

The Bovie
Smoke Shark Surgical Smoke Evacuator is intended for use in the electromagnetic
®
environment specified below. The customer or user of the Bovie
IEC 60601
Immunity Test
Test Level
Electromagnetic
+6 kV contact
discharge (ESD)
IEC 61000-4-2
+8 kV air
Electrical fast
+2 kV for power
transient/burst
supply lines
IEC 61000-4-4
+1 kV for input/
output lines
Surge
+1 kV differential
mode
+2 kV common
IEC 61000-4-5
mode
Voltage dips, short
<5 % U
interruptions, and
(>95 % dip in U
voltage variations
for 0.5 cycle
on power supply
input lines.
40 % U
(60 % dip in U
IEC 61000-4-11
for 5 cycles
70 % U
(30 % dip in U
for 25 cycles
<5 % U
(>95 % dip in U
for 5 sec
Power frequency
3 A/m
(50/60 Hz)
magnetic field
IEC 61000-4-8
Customer/Technical Services: 800.251.3000 (United States) or +1 615.964.5532 (International)
OPERATING INSTRUCTIONS: SECTION 2.0
Table 2
should ensure that it is used in such an environment.
Compliance
Level
+6 kV contact
+8 kV air
+2 kV for power
supply lines
+1 kV for input/
output lines
+1 kV differential
mode
+2 kV common
mode
<5 % U
T
T
)
(>95 % dip in U
T
for 0.5 cycle
40 % U
T
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
T
)
(>95 % dip in U
T
for 5 sec
3 A/m
Smoke Shark Surgical Smoke Evacuator
®
Electromagnetic Environment - Guidance
Floors should be wood, concrete or ceramic
tile. If floors are covered with synthetic
material, the relative humidity should be at
least 30 %.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
Mains power quality should be that of a
typical commercial or hospital environment.
)
T
If the user of the Bovie
Surgical Smoke Evacuator requires
continued operation during power main
interruptions, it is recommended the Bovie
)
Smoke Shark Surgical Smoke Evacuator be
T
powered from an uniterruptible power supply
or a battery.
)
T
)
T
Power frequency magnetic fields should be
at levels characteristic of a typical location in
a typical commercial or hospital environment.
P000027200 Rev A Bovie
Smoke Shark
®
®
Page 14
Smoke Shark III Operator's Manual
®