Guidance and Manufacturer's Declaration - Electromagnetic Emissions
The Bovie
Smoke Shark Surgical Smoke Evacuator is intended for use in the electromagnetic
®
environment specified below. The customer or user of the Bovie
IEC 60601
Immunity Test
Test Level
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to
2.5 GHz
Conducted RF
IEC 61000-4-6
150 kHz to
80 MHz
NOTE 1: At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE 2: These guidelines may not apply in all situations. Electromagnetic propagation is affected by
absorption and reflection from structure, objects and people.
Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and
land mobile radios, amateur radio, AM and FM radio broadcast and TV broadcast cannot be predicted
theoretically with accuracy.
To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey
should be considered. If the measured field strength in the location in which the Bovie
Surgical Smoke Evacuator is used exceeds the applicable RF compliance level above, the Bovie
Shark Surgical Smoke Evacuator should be observed to verify normal operation.
If abnormal performance is observed, additional measures may be necessary, such as reorienting or
relocating the Bovie
Smoke Shark Surgical Smoke Evacuator.
®
Over the frequency range 150 kHz to 80 MHz, field strengths should be less than 3 V/m.
Customer/Technical Services: 800.251.3000 (United States) or +1 615.964.5532 (International)
OPERATING INSTRUCTIONS: SECTION 2.0
Table 3
should ensure that it is used in such an environment.
Compliance
Level
3 V/m
3 Vrms
Smoke Shark Surgical Smoke Evacuator
®
Electromagnetic Environment - Guidance
Portable and mobile RF communications
equipment should be used no closer to any
part of the Bovie
Smoke Shark Surgical
®
Smoke Evacuator including cables, than the
recommended separation distance calculated
from the equation applicable to the frequency of
the transmitter.
d = 1.7 √P 80 MHz to 800 MHz
d = 2.3 √P 800 MHz to 2.5 GHz
d = [3.5/V1} √P
Where P is the maximum output power rating
of the transmitter in Watts (W) according to
the transmitter manufacturer and D is the
recommended separation distance in meters (M).
Field strength from fixed RF transmitters, as
determined by an electromagnetic site survey,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of
equipment marked with the following symbol:
P000027200 Rev A Bovie
®
Smoke Shark
®
Smoke
®
Page 15
Smoke Shark III Operator's Manual