3 Performance specifications
Performance characteristics of sterilization programs
The results in this table show which inspections were performed on the steam sterilizer. The marked fields demonstrate
compliance with all the applicable sections of the standard
Type tests
Program type in
accordance with
EN
}
13060
Dynamic pressure test
}
the sterilization chamber
Air leakage
}
Empty chamber test
}
Solid load
}
Porous partial load
}
Porous full load
}
Simple hollow body
}
(}
Hollow body
B)
Product with narrow
}
lumen (}
Hollow body
A)
Single wrapping
}
Multiple wrapping
}
Drying
solid
load
}
Drying, porous load
Sterilization temperature
Sterilization pressure
Sterilization time
X = Complies with all applicable sections of the standard
10
Universal-B
Type B
of
X
X
X
X
X
X
X
X
X
X
X
X
134°C
2.1 bar
05:30 min
EN
13060.
}
Quick-S
Gentle-B
Type S
Type B
X
X
X
X
X
X
X
X
--
X
--
X
X
X
--
X
--
X
--
X
X
X
--
X
134°C
121°C
2.1 bar
1.1 bar
03:30 min
20:30 min
EN 13060
}
Prion-B
Type B
X
X
X
X
X
X
X
X
X
X
X
X
134°C
2.1 bar
20:30 min