Event
Possible cause
10063
The manual feed water supply is activated.
The device must be filled with at least 1.5 l of
demineralised water.
10082
In the program start, the counter for the feed
quantity is evaluated. The limit value is
exceeded. Therefore, starting the program is
not possible.
10093
In the program start, the counter for the rinse
value is evaluated. The limit value is
exceeded. Therefore, starting the program is
not possible.
10094
In the program start, the system checks
whether the vacuum test can be performed
successfully at the current ambient
temperature.
10098
A failure of the supply voltage was detected
in the program run.
10099
A failure of the supply voltage was detected
in the program run of a service program.
10101
The float switch (S13) in the overflow funnel
detects a short-term impermissible water
level, which indicates a blockage in the
wastewater system.
What you can do
Please supply the device with sufficient
demineralised water before starting the
program or ensure an automatic water
supply via a water treatment unit.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
The automatic emptying was skipped
several times because the drying was ended
manually. Starting the program is not
possible before the "Empty double jacket"
service program has been carried out.
Please start the "Empty double
jacket"service program.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
The automatic emptying was skipped
several times because the drying was ended
manually. Starting the program is not
possible before the "Empty double jacket"
service program has been carried out.
Please start the "Empty double
jacket"service program.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
The ambient temperature of the device is
very high.
Allow the device to cool down.
Observe the set-up conditions.
Ensure that the device is sufficiently
ventilated.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
Connect the device to a specially fused
power supply to which no other electrical
device is connected.
Check the power cable on the rear of the
device to ensure that it is firmly seated.
Attach the safety latch.
Connect the device to a specially fused
power supply to which no other electrical
device is connected.
Check the power cable on the rear of the
device to ensure that it is firmly seated.
Attach the safety latch.
Short-term wastewater blockage - please
check wastewater hose for kinks or, if
applicable, a closed shut-off.
This is only a warning. The reprocessing
result is not affected. You can continue to
use the device.
14 Malfunctions