14 Malfunctions

14 Malfunctions
Not all notifications on the display are malfunction messages. Warnings and malfunction messages are issued on the
display with an event number. This number serves identification purposes.
Nature of the display noti-
fication
Messages
Warnings
Malfunction messages
Troubleshooting online
All messages with current descriptions can be found in the Troubleshooting portal
on the MELAG website (https://www.melag.com/en/service/troubleshooting).
Before contacting customer services
Ensure that you have complied with all instructions relating to a warning or malfunction message issued on the display of
the device. The following table contains a summary of the most important events. Should you be unable to find the rele-
vant event, or your efforts do not redress the problem, you can contact your nearest stockist or an authorised customer
service provider. To enable us to give the best possible service, please have the serial number of your device, the mal-
function number and a detailed description of the malfunction to hand.

Malfunction logs

In the Logs > Malfunction logs menu, you can view malfunction logs and output them to a USB stick.
See also:
Logs menu
[} page 35]
Log list
[} page 36]

Warning and malfunction messages

Event
Possible cause
10059
The external wastewater container is full.
10062
The lack of water in the feed water tank
could not be remedied within the monitoring
time.
58
Description
Many messages are notifications. They serve to inform and provide
assistance in the operation of the device.
Warning messages are displayed when necessary. Warning messages
contain instructions that help to ensure smooth operation and to recognise
undesirable conditions. Comply with these warnings early in order to avoid
malfunctions.
Malfunction messages are issued when it is not possible to ensure safe
operation or safety of sterilization. These can appear on the display shortly
after activating the steam sterilizer or during a program run. If a malfunction
occurs during a program run, the program will be aborted.
What you can do
Please empty the external wastewater
container before the next program start.
Ensure water supply (main valve) or fill in
case of supply from canisters