5.3.2
Jet Polisher
WARNING: Contraindications. Patients
on a sodium restricted diet must not receive
treatment with sodium bicarbonate powders.
WARNING: Contraindications.
Patients who suffer from serious respiratory
problems such as chronic bronchitis,
asthma, emphysema, etc., must not receive
prophylaxis treatment unless otherwise
indicated by their doctor.
WARNING: Contraindications. Patients
wearing contact lenses or glasses must
remove them prior to receiving treatment
with the jet polisher.
WARNING: Contraindications - Jet
polisher. Do not aim the air, supragingival
powder, or water jet onto the soft tissues or
inside the gingival sulcus. Failure to comply
with this prescription can cause a gingival
tissue emphysema (emphysema of the
mucous and/or subcutaneous). For this type
of application, use only subgingival powder.
5.4
Instructions for Use - Scaler Part
After having connected all the accessories as described in
as follows:
Lift the scaler handpiece, with or without
insert, fill the irrigation circuit by selecting
"flush" on the keyboard.
The "flush" function LED flashes.
The appliance allows the use of two types
of irrigation: the normal water circuit or
the tank circuit.
CAUTION: Make sure the polisher
handpiece is stowed in its housing,
otherwise the system will remain
inactive.
32
! WARNING: Contraindications. Do
not use the device near areas subject to
recent dental extraction and in traumatised/
damaged areas (or areas nearby) due to the
risk of emphysema.
WARNING: Temperature of the water
spray. The appliance is equipped with
a double safety device that controls the
temperature of the water spray. Before the
treatment, it is however recommended to
instruct the patient to inform the operator
if he perceives an excessive increase in the
temperature of the water.
WARNING: Infection control and
cleaning the water and air circuits. To
ensure utmost patient and operator safety,
after each treatment follow the indications
provided in
Chapter 8 on page 49
WARNING: Do not use the appliance
without water. Make sure that the appliance
is connected to the water circuit and the
water tap is open.
CAUTION: Do not attempt to unscrew the
tap of the powder container before having
performed the "refill" cycle.
Chapter 4.3 on page 18
1
.
proceed
refill