combi touch
Check the integrity of the filter, rinse it,
dry it with compressed air, making sure to
remove any residual impurities;
Completely immerse the male coupling
for the external water circuit connection
in the disinfectant solution (CIDEX
OPA). Leave to soak for 12 minutes at
20 °C ±2 °C.
Remove the male coupling for the
external water circuit connection from
the disinfectant solution (CIDEX
and rinse it under running water;
Place the male coupling for the external
water circuit connection in a clean and
safe place for later use.
Disinfect the internal surfaces of the
seat where the male coupling for the
external water circuit connection will be
inserted using a clean, soft, low-lint cloth
dampened with the disinfectant solution
(CIDEX
OPA).
®
52
4
5
®
6
OPA)
®
7