Press the "flush" button on the touch
keyboard (see
Chapter 6.2 on page 43
Select the external irrigation circuit ("tap"
Chapter 6.1 on page 40
symbol - see
Press and release the pedal. When the
"flush" cycle begins, start injecting the
distilled water with the syringe. The
"flush" cycle on the "Ultrasound" part
lasts about 23 seconds;
Repeat points 21 to 25 another 4 times;
Repeat points 14 to 18 twice until the
irrigation circuit is completely drained of
the water used in the previous steps.
NOTE: It is advisable to drain the circuit
during long periods of non-use. If the
device is used after the disinfection
procedure, and provided different liquids
don't need to be used for irrigation,
the water circuit does not need to be
drained.
CLEANING AND STERILISATION
23
);
24
);
25
26
27
refill
refill
67