Press and release the pedal. When the
"flush" cycle begins, start injecting the
disinfectant with the syringe.
During the "flush" cycle, inject 50ml
of disinfectant, leaving the remaining
10ml in the syringe.
The "flush" cycle on the "Polishing" part
lasts about 40 seconds;
NOTE: The "flush" cycle on the
"Polishing" part cannot be interrupted.
Before running the "flush" cycle on the
"Polishing" part check that the right-
hand knob in the front part of the device
(Ref. 9 inside cover) is open. At the end
of the "flush" cycle, stop injecting the
liquid with the syringe.
CAUTION: Do not press too hard on
the piston of the syringe. The irrigation
circuit hoses are quite small and it is
normal for injection to occur slowly.
NOTE: If the 50ml of disinfectant
are injected before the flush cycle is
completed, wait for it to finish. If the
cycle ends before the full 50ml have
been injected, repeat the flush cycle until
the 50ml have been completely injected.
Wait 12 minutes;
CLEANING AND STERILISATION
16
17
55