DENTAL LED OPERATING LIGHT POLARIS
accordance with its regulations with a power of 10
A–250 V and that in series during the secondary
windings would be put 2 rapid fuses 2.5 A.
!
ATTENTION
Device must be installed only by authorized techni-
cians.
Installation and all kinds of maintenance operations
must be realized when the light is switched off.
AUXILIARY POWER SUPPLY OUTPUT
An auxiliary power supply connector is placed in the
back of light head body. The auxiliary power supply
connector has to be used only to supply G.COMM
devices (for example the G.COMM bleaching dental
light COREWHITE).
Maximum power supply provided:
Maximum power provided:
INSTALLATION
The operating light "POLARIS" is delivered as one
LIGHT HEAD and one ARM packed separately.
The complete installation of the light requires the as-
sembly of these two parts.
PACKAGING:
470
Proceed as follows:
1. Take the arm out of the box.
2. Use an eyelet to connect the ground cable at the
bottom of the pivot (M3 thread). See Figure 3.
3. Connect the cable of the arm with the cable that
comes out from the dental unit column. Place the
arm with mounted pivot in the bushing of the col-
umn. Check the movements of the dental operating
Figure 3
connection for protective
earth conductor
17-24 Vac
3 A
bushing
unit column
Connector
green 0V
black 17-24V
Figure 2A - LIGHT HEAD
white in not
connected
Figure 2b - ARM
light and the stability of the entire structure (see
Figure 3).
4. Carefully take the "POLARIS" light head out of its
protective package in order to avoid damage to glass
protection or other parts.
5. As per Figure 4 connect the male connector placed
in the arm to the female connector placed in the light
head. Insert the light head pivot in the bend (called
light head bend); make sure that the threaded hole of
the pivot is in line with the small one on the arm.
Place the M5 screw supplied with the light into the
hole and screw it all the way until locking position.
Check for stability.