15.3 Cleaning
• Instrument
The instrument is cleaned with a wet cloth. Any
of the above disinfectants can be used. For use in
environments requiring sterility, the instrument should
be covered with a sterile cover.
• MulTipeg Driver and MulTipegs
The MulTipeg Driver and the MulTipeg should be
cleaned with water with or without detergent, using
a light brush. Any of the above disinfectant fluids can
be used. For use in environments requiring sterility, the
MulTipeg Driver and MulTipeg should be autoclaved
before use according to the above instructions.
16. Lifetime
The lifetime of the instrument is determined by the
number of charging cycles. The internal batteries can be
fully charged more than 500 times before they need to be
replaced. The instrument should not be left uncharged for
more than 1 year.
The MulTipeg Driver is guaranteed for at least 100
autoclave cycles, and a MulTipeg is guaranteed for at least
20 autoclave cycles, before they are degraded in any way.
17. Troubleshooting
17.1 Possible errors
• Difficult to achieve a measurement:
In some cases it is more difficult for the instrument to
bring the MulTipeg into vibration. If so, try to hold the
instrument tip closer to the tip of the MulTipeg. Check
also that no soft tissue is touching the MulTipeg which
might stop its vibration.
• Noise warning (audible and visible on the display):
An electric device close to the instrument is causing the
warning. Try to remove the source.
• The instrument suddenly turns off:
The instrument turns off automatically after 30 seconds
of inactivity. It also turns off if the battery level is too
low, and due to any of the error codes described below.
• Not all segments are lit up when instrument is started:
The instrument is damaged and has to be sent for repair
or exchange.
17.2 Error codes
If malfunctioning, these error codes are shown on the
display before it turns off:
E1: Hardware error. Malfunctioning electronics
E2: Noise error. Shown if constant electromagnetic noise
is present
E3: Pulse power error. Malfunctioning magnetic pulse
generation
Use of accessories other than those specified or
provided by the manufacturer of this equipment
could result in increased emissions or decreased
electromagnetic immunity of this equipment and
result in improper operation.
18. Accessories & Spare Parts
Model
MulTipeg
Sterile
Driver
Cover
REF
55003
55105
Model
EU plug
UK plug
REF
55094
55095
55264
55265
MulTipeg: Please refer to the updated list from the
supplier.
19. Repair
In case of a malfunctioning instrument, contact the
manufacturer or distributor. Osseo 100 is covered by a
two-year warranty.
5
English
Mains adapter
Model No.
UE05WCP-052080SPC
Or
UES06WNCP-052080SPA
55093
55263
AU plug
US plug ISQ tester
55096
55097
55217
55266
55266