6.6 Adjusting the Humeral Rotation Feature
6.7 Installing the Pull Cable
CAUTION
Removing the pull cable
Injury due to malfunction of the product.
► The pull cable on the 12K501 AxonArm Ergo is to be used for emergency operation only. For safety reasons, it
must not be removed. When the Axon-Bus prosthethic system is switched off or defective and when the bat
tery is empty, the AxonArm cannot be locked or unlocked without the pull cable.
6.8 Checking the Symmetry
1) Connect the Axon-Bus gripping component (e.g. Michelangelo Hand) to the AxonArm.
2) Check the symmetry with respect to the other hand (e.g. using the 743L20=230 Ottobock Laser-Line).
7 Handling
7.1 Charging the Battery
The 757B501 AxonEnergy Integral must be charged prior to initial use. The 757B501 AxonEnergy Integral may
only be charged with the 757L500 AxonCharge Integral. The charging process can only be performed with the
13E500 AxonMaster connected.
1) Connect the charging plug of the 757L500 AxonCharge Integral to the charging receptacle.
→ The beeper sounds briefly 2x.
→ The Axon-Bus prosthethic system is deactivated and the charging process starts automatically.
2) To activate the Axon-Bus prosthethic system, disconnect the charging plug and switch on the Axon-Bus pros
thethic system.
INFORMATION: The prosthesis cannot be used during the charging process.
7.2 Switching On the Axon-Bus Prosthetic System
► Press the charging receptacle button and hold for one second (Fig. 2, arrow).
→ The beeper emits two short signals.
7.3 Switching Off the Axon-Bus Prosthetic System
► Press the charging receptacle button and hold for more than one second (Fig. 2, arrow).
→ The beeper emits one long signal.
7.4 Determining the Battery Charge Level
The battery charge level can be queried at any time.
AxonArm Hybrid 12K500=*, AxonArm Ergo 12K501=*
► The humeral rotation turntable has integral stops
(+/- 80 degrees) to block extreme movements.
Turntable friction is easily adjusted by turning of the
external adjustment screw, using a coin or similar
object.
1) Bring the forearm into neutral position (no outward
or inward rotation).
2) Affix the cable guide to the socket with cap screws
(centred in relation to the strap clamp and approx.
30 mm away from the lamination ring).
3) Test inward and outward rotation of the socket up to
the stop. Ensure that the pull cable cannot be ten
sioned (loop).
Handling
Ottobock | 33