Preparation for Use
6.1 Installing the AxonRotation Adapter
6.2 AxonMaster
Various control modes and switching methods as well as the option of implementing a fine-tuned patient-specific
adjustment are available for optimum adaptation to the patient. For information please refer to the 647G590
Instructions for Use for the 13E500 AxonMaster.
Two MyoBock electrodes, e.g., 13E200/202=*, or if necessary a pull switch, e.g., 9X18, can be connected to the
13E500 AxonMaster.
Input on AxonMaster
Input 1
Input 2
Input 3
► Remove the cover caps from the required inputs. The remaining inputs cannot be used at this time and there
fore remain covered by the caps.
6.2.1 Controlling the Electronic Lock
The electronic lock can be controlled myoelectrically by the following means configured at the factory:
•
A long co-contraction on channels 1 and 2 of the AxonMaster releases or locks the lock.
•
A short co-contraction switches the gripping mode of the gripper.
•
A pulse on channel 3 of the AxonMaster releases or locks the lock.
28 | Ottobock
1) Use silicone grease to seal the plug connection.
2) Connect the AxonRotation Adapter with the con
necting cable.
3) Position the AxonRotation Adapter such that the
middle screw holes on the lamination ring can
be used.
INFORMATION: The other screw positions can
be used for correcting the alignment of the
AxonRotation Adapter after the lamination ring
has been glued in.
4) Screw the AxonRotation Adapter to the lamination
ring.
Electrode
Hand closing
Hand opening
Pull switch (optional)
AxonArm Hybrid 12K500=*, AxonArm Ergo 12K501=*