Donning The Prosthetic Glove; Legal Information - Ottobock Michelangelo Hand Transcarpal 8E550 Instrucciones De Uso

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 50
1) (see fig. 8) Loosen the 5 screws in the connection plate slightly and twist the hand. A correc­
tion of ± 15 ° is possible.
2) Then tighten the screws to a torque of 1.5 Nm.
3) (see fig. 9) Apply silicone grease to the connector plugs and connect them.
→ Position the cables and plugs such that they will not be damaged when putting on the
Michelangelo Hand.
4) Place the hand on the socket.
5) (see fig. 10) Insert the provided screws into the bore holes on the sides and tighten them to
1.5 Nm.

6.2 Donning the prosthetic glove

For daily use, the Michelangelo Hand must be worn with the AxonSkin prosthetic glove. The glove
protects the mechanism against environmental influences such as moisture, dirt and dust.
For information about the use (putting on and taking off) and care of the prosthetic glove, please
refer to the instructions for use included with the prosthetic glove.
6.3 Settings
Adjustments to the product can be made via Bluetooth data transmission using the 560X500=*
AxonSoft software. For this purpose, a wireless connection must be established between the
13E500 AxonMaster and the PC with the help of the 60X5 BionicLink PC.
For further information, see the instructions for use provided with the 13E500 AxonMaster and the
560X500=* AxonSoft software.
7 Cleaning
NOTICE
Improper product care
Damage to the product due to the use of incorrect cleaning agents.
Only clean the product with a damp cloth and mild soap (e.g. 453H10=1 Ottobock Der­
maClean).
1) Clean the product with a damp cloth and mild soap (e.g. Ottobock 453H10=1 Derma Clean)
when needed.
Ensure that no liquid penetrates into the system component(s).
2) Dry the product with a lint-free cloth and allow it to air dry fully.
8 Maintenance
► Inspect the entire prosthesis for wear during normal consultations.
8.1 Service
Since all moving mechanical components are subject to wear and tear, regular service is required
within the guarantee period. The entire product is inspected by Ottobock Service (Ottobock Myo-
Service). Wear parts are replaced as required.
Failure to adhere to the service schedule invalidates the warranty.

9 Legal information

All legal conditions are subject to the respective national laws of the country of use and may vary
accordingly.
9.1 Liability
The manufacturer will only assume liability if the product is used in accordance with the descrip­
tions and instructions provided in this document. The manufacturer will not assume liability for
22