Participation in an Ottobock product training course is mandatory prior to using the product.
During this product training course you will receive a password giving you access to the
adjustment software. Additional product training courses may become necessary to qualify
for software updates.
Do not share your unlock PIN.
Use the online help which is integrated into the software.
CAUTION
Incorrect electrode settings
Injury due to unexpected product behaviour.
The electrodes are to be placed on intact skin only and with as much electrode-skin contact
as possible. In the case of strong interference from electronic devices, the position of the
electrodes should be checked and changed if necessary. If the interference cannot be elim­
inated or if you do not achieve the expected results by adjustment or selection of the appro­
priate control programme, please contact the Ottobock branch responsible for your country.
Set the electrode gain as low as possible in order to reduce interference from powerful elec­
tromagnetic radiation (e.g. visible or or concealed theft prevention systems at the
entrance/exit of stores), metal detectors/body scanners for persons (e.g. in airports) or other
sources of strong electromagnetic interference (e.g. high-voltage lines, transmitters, trans­
former stations, computer tomographs, magnetic resonance tomographs, etc.).
INFORMATION
Do not use silicone spray when putting on the prosthetic glove. This impairs the secure fit of the
glove.
Observe the instructions for use provided for the prosthetic glove and the process for putting it
on and taking it off described in the same.

4.5 Information on Use

CAUTION
Mechanical stress on the product
Injury due to faulty control or malfunction of the product.
Do not subject the product to mechanical vibrations or impacts.
Check the product for visible damage before each use.
CAUTION
Operation outside of the permissible temperature range
Injury due to faulty control or malfunction of the product.
Avoid operation in areas outside the permissible temperature range (see Page 24).
CAUTION
Penetration of dirt and humidity into the product
Injury due to unexpected product behaviour of the prosthesis system.
Ensure that neither solid particles nor liquids can penetrate into the product.