TROUBLESHOOTING

Type of problem
Water comes into the powder chamber or
leaks from the cap
No powder/air jet coming from the unit
Air and/or powder is leaking from the
threads of the cap
7KH XQLW HI¿FLHQF\ GLPLQLVKHV
56
Solutions
Check the dental unit connection
Check the state of the O-rings of the turbine connection
Clean the cap and the powder chamber
Clean the handpiece
&RQQHFW WKH XQLW WR WKH GHQWDO XQLW EH FDUHIXO WKH SRZGHU FKDPEHU PXVW EH HPSW\
&KHFN WKH TXDOLW\ RI WKH VSUD\
Fill the powder chamber
Send the unit to an approved EMS repair center
Cut immediately the air input releasing the foot pedal
Wait for 1-2 minutes for the system to decompress
Disconnect the unit from the dental unit
Press the foot control pedal of the dental unit
,I QR DLU FRPHV RXW IRU WKH WXUELQH FRQQHFWRU LW LV \RXU GHQWDO XQLW WKDW FDXVHV WKH SUREOHP
,I DLU FRPHV RXW LW LV WKH GHYLFH WKDW FDXVHV WKH SUREOHP
8QVFUHZ WKH FDS ZLWK WKH WRS RI WKH ZDVKKDQG EDVLQ EXV RI WKH SRZGHU FDQ EH HMHFWHG $ VHDOHG DSSDUDWXV HYHQ GLVFRQQHFWHG
can indeed remain under pressure
Empty the powder chamber and screw the cap back on
&RQQHFW WKH XQLW WR WKH GHQWDO XQLW EH FDUHIXO WKH SRZGHU FKDPEHU PXVW EH HPSW\
Connect the handpiece to the unit
Press the foot control pedal of the dental unit
,I DLU FRPHV WKURXJK WKH KDQGSLHFH FRQQHFWRU WKH KDQGSLHFH LV EORFNHG &OHDQ WKH KDQGSLHFH
,I QR DLU FRPHV RXW RI WKH KDQGSLHFH FRQQHFWRU WKH XQLW LV EORFNHG
Send the unit to an approved EMS repair center
Check the seal and the cleanliness of the thread on the powder chamber and on the cap
Replace the seal if necessary
$ SRZGHU UH¿OO LV SHUKDSV QHFHVVDU\
Clean the handpiece
Send the unit to an approved EMS repair center