Ambient conditions
Storage and transport without packaging
Operation
General information
Reference number
Weight of the bebionic
hand small
Weight of the bebionic
hand medium
Opening width
(between index finger
and opposed thumb)
Extension/flexion of the
wrist
Service life
Behaviour of the ter
minal device during
charging
Version of the terminal
device
Load limits
Force on single finger (static)
Transverse force on single finger (static)
Force on chassis (static, supporting the hand)
Force with closed hand (static, carrying a bag) 152 N
Forces on thumb (static)
Data transfer
Wireless technology
Range
Frequency range
Modulation
Maximum output power
11.1 Torque values of the screw connections
Screw connection
Adjustment screw for the mediolateral thumb
position (see Page 46)
56
bebionic hand EQD
8E70=*
433 g/0.95 lbs
616 g/1.36 lbs
–
The terminal device has no function
Hardware and firmware version accessible via the adjustment software
-25°C/-13°F to +70°C/+158°F
Max. 90% relative humidity, non-condensing
-5°C/+23°F to +45°C/+113°F
Max. 95% relative humidity, non-condensing
bebionic hand Short
Wrist
8E71=*
Approx. 402 g/0.89
lbs
Approx. 588 g/1.3 lbs
75 mm
–
5 years
32 N
44 N
500 N
40 N
Bluetooth Smart/Low Energy
min. 2 m / 6.7 ft
2402 MHz to 2480 MHz
GFSK
9.6 dBm
Tightening torque
bebionic hand small
2 Nm/18 lbf. In.
bebionic hand Flex
8E72=*
Approx. 504 g/1.1 lbs
Approx. 689 g/1.52
lbs
-40° to +40° in 20°
increments
bebionic hand medium
1 Nm/9 lbf. In.