Fluid Bag
Fluid Bags (not included) hang on the IV Hooks at the top of the Fluent Fluid Management System. Hooks provide the ability to hang up to 6 liters
of fluid containing hypotonic, isotonic, ionic and non-ionic distention fluids at once.
The Fluid Bag Tube connects to the blue Inflow Tube Set, which connects to the hysteroscope Inflow Channel Tube. Fluid is pulled from the Fluid
Bag through the Inflow Tube Set and delivered through the hysteroscope Inflow Channel Tube into the patient's uterus.
WARNING!
Placing excessive force or weight on the IV Hooks can overload the scale connected to the IV Hooks. Doing so may result in an inaccurate fluid
deficit value, causing risk to patient safety.
WARNING!
When performing monopolar hysteroscopic electrosurgery, the distension medium must be electrically non-conductive. Examples include glycine,
sorbitol and mannitol. Isotonic irrigation fluids may only be used when performing bipolar electrosurgical resection procedures. Examples include
saline and lactated Ringer's solution.
Touchscreen Monitor
The Touchscreen Monitor includes two system speakers and the Touchscreen User Interface. Tilt and swivel the monitor for optimal viewing.
Caution!
Only use the Handle to move or position the Fluent Fluid Management System. Do not pull or push the system by the Touchscreen Monitor.
Touchscreen User Interface
Use the Touchscreen User Interface to configure and view system information, set the deficit limit, prime the system, and make other adjustments
as needed. The Touchscreen prompts performing a task (such as Hang Fluid Bag) and displays task status (such as Fluid Bag Hung).
Touchscreen Icons
Icon
Name
Help
System Settings
Clear
Back
Next
Prime
Confirm
Pause
Change
Down/Up
Accept/Confirm
English
Fluent Fluid Management System
Description
Displays additional information
Displays the System Settings screen
Clear
Clears an error condition
Displays the next step or screen
Primes the system
Confirms the system request
Pauses the procedure. When the system is paused, this button changes to Run to
restart the procedure.
Displays the available settings for the selected item
Decreases or increases the settings for the selected item
Applies the change to the selected setting
Chapter 1: Introduction to Fluent Fluid Management System
13