4.
Pull out the fuse holder and check the fuse.
5.
Insert the new fuse. Use only the type of fuse: T5AH, 250 V fuses.
6.
Insert the fuse holder.
7.
Reconnect the Power Cord to the wall outlet and turn the system On (|) to ensure it is working properly.
Load Cell Calibration Check
The fluid deficit measurement test checks the fluid bag load cell and the waste bag load cell and the accurate measurement of weight to ensure
that all elements are functioning properly. This test requires a 500 gram weight which is shipped with the system.
1.
Turn the system ON.
2.
Touch button ( ) for settings screen
3.
Hang 500 gram weight on one of the fluid bag hooks
4.
Touch the button for 'Load Cell Calibration Check'
5.
Number displayed for the 'Supply Weight' should be 500 ±25g
6.
Remove weight from the fluid bag hook and hang the 500 gram weight on one of the waste bag hooks
7.
Number displayed for 'Collection Weight' should be 500 ±25g
8.
Remove weight from the waste bag hook
9.
Select 'Exit' to return to the setup screen.
Pressure Calibration Check
The pressure measuring test checks the pressure accuracy to ensure that all elements are functioning properly. This test requires an in-FloPak,
3000mL fluid bag and an under buttocks drape or bucket. The fluid bag is placed on the supply load cell hanger to produce the hydrostatic
pressure that is used to test the pressure sensor.
1.
Turn the system ON.
2.
Place the in-FloPak onto the console.
3.
Touch ( ) button for settings screen
4.
Choose 'Pressure calibration check'
5.
Choose 'Lock'
6.
Hang the 3000mL of fluid bag on one fluid bag hook. Make sure that bag clamp is pinched. (Fluid will spill out if not done)
7.
Fully pinch both inflow clamps
8.
Remove bag spike cap and spike bag with one inflow spike, leave other spike capped.
9.
Place a bucket onto the ground and place the Hysteroscope Inflow Channel Luer lock into the bucket.
10.
Open clamp connected to full bag
11.
Open clamp connected to the in-FloPak
12.
Open clamp connected to the Hysteroscope Inflow Channel Luer Lock
13.
Touch 'Start' to fill tube with fluid. When tube is filled and air bubbles in line are gone, touch 'Stop'.
14.
Remove the Luer Lock from the bucket and hold it even with top of fluid bag hanger.
15.
Apply firm, even pressure on the In-FloPak to ensure it is seated properly.
16.
Number displayed for pressure reading should be 20±5 inH20.
17.
Place the Luer Lock back into the bucket.
18.
Select 'Exit' to return to the setup screen.
19.
Pinch fluid bag clamp.
20.
Remove bag spike from the fluid bag. Then remove cartridge from the console to allow fluid to drain into the bucket.
English
Fluent Fluid Management System
Chapter 8: Maintenance
33