Cancel/Decline
Run
Complete Procedure
Exit
New Procedure
Last Procedure
End
Zero
Reprime
Done
Fluent Procedure Kit
The Fluent Procedure Kit contains the Fluent In-FloPak, the Fluent Out-FloPak, the Waste Bag, and the Tissue
Trap.
WARNING!
Do not reprocess of sterile disposable products
Reuse of Fluent In-FloPaks or Fluent Out-FloPaks can cause an infection hazard for patients and/or users as
well as impair product functionality. Contamination and/or impaired functionality of the system can cause risk of
injury, illness, or death. Do not re-process any of the components contained within the single-use Procedure Kit.
Fluent In-FloPak
The blue Fluent In-FloPak pulls clean fluid from the Fluid Bag.
The Fluent In-FloPak contains the Fluid Bag Tube and the hysteroscope Inflow Tube. It fits securely into the blue
Fluent In-FloPak Receptacle on the front left side of the Fluid Management System.
Before each procedure, connect the tubes as follows:
• From the Fluid Bag Tube spike to the Fluid Bag
• From the hysteroscope Inflow Tube (that has a blue band adjacent to the Luer lock connector) to the Hysteroscope Inflow Channel
These connections allow the transfer of irrigation fluid from a Fluid Bag to the hysteroscope Inflow Channel. The
flow of fluid is monitored and controlled using the Touchscreen to maintain the pressure at a specified setting.
Fluent Out-FloPak
The yellow Fluent Out-FloPak drains waste fluid from the hysteroscope Outflow Channel, the MyoSure Tissue
Removal Device (TRD) Tube, and the Under Buttocks (UB) Drape Tube into the Waste Bag.
The Fluent Out-FloPak contains the Waste Bag Tube (which contains the in-line Tissue Trap), the hysteroscope
Outflow Tube, the MyoSure Tissue Removal Device (TRD) Tube, and the Under Buttocks (UB) Drape Tube. It fits
securely into the yellow Fluent Out-FloPak Receptacle on the front right side of the Fluid Management System.
Before each procedure, connect the tubes as follows:
• From the Waste Tube connector to the Waste Bag
• From the hysteroscope Outflow Tube (that has a yellow band adjacent to the Luer lock connector) to the
Hysteroscope Outflow Channel
14
Fluent Fluid Management System
Cancels the change to the selected setting
Starts the pumps running to provide irrigation. When running, this button changes to
Pause to halt the pumps.
When ending a procedure, the Outflow pump continues to run to collect the
remaining flow. Pressing this button stops the Outflow pump and displays the values
for Total Deficit, Total Fluid Volume, Final Pressure, and Cutting Time.
Returns to previous screen
Displays the System Setup screen to start a new procedure
Displays results from the previous procedure
Ends the procedure and begins evacuating the Out-FloPak tubing
Resets the procedure's deficit value to zero
Primes the system again after a procedure has started
Ends evacuation of the Out-FloPak tubing
Chapter 1: Introduction to Fluent Fluid Management System
Fluent In-FloPak
Fluent Out-FloPak
English