Error Text
Check waste bag hooks-
Deficit limit exceeded -
Pressure too high -
Priming error, ensure scope
end not blocked -
English
Fluent Fluid Management System
Error Code
Description
02010027
Check if the hooks were handled, or
if the waste bag has been excessively
jostled.
This may also occur when excessive
force is momentary imparted on the
hooks during a bag change or when
moving the console.
02010028
Deficit limit exceeded
020A0005
Measured uterine pressure is greater
than 165 mmHg for over 5s.
End of scope may be up against
uterine fundus.
020A0007
The system has detected pressure
buildup during priming.
Chapter 10: Troubleshooting
Resolution
Keep hands off of waste hooks.
Remove any object other than the bag
from hooks.
If bag is moving, stabilize it, wait a
few seconds and touch 'Clear'
End the procedure.
If the doctor determines that it is
clinically safe, increase the deficit
limit, touch Clear, and resume running
the procedure.
Ensure scope is clean and tip is not
blocked.
Ensure the inflow stopcock and
outflow stopcock are both open.
During priming, the scope must be
open to the air at the level of the
patient.
Ensure that the fluid flow at the end
of the scope is not blocked.
Scope must not be in the uterus when
priming.
Ensure that inflow tube clamp and
scope stopcock are open to allow
fluid to flow freely. All scopes must be
primed prior to use.
To continue procedure, exit this
screen:
1. Touch 'Clear'
2. Remove scope from uterus
3. Reprime the scope before continu-
ing the procedure.
39