3487ALead_CH.fm 5/13/04 3:14 pm
UC200xxxxxx EN
4 x 8 inches (101 mm x 203 mm)
Table 1. Device specifications
Description
Connector
Shape
b,c
Conductor resistance
Length
Diameter
Distal end
Number of electrodes
Electrode shape
Electrode length
Electrode spacing
Electrode distance
Distal tip distance
Radiopaque marker
Proximal end
Lead contact length
Lead contact distance
Stylet handle length
a
All measurements are approximate.
b
Electrical resistance of this device only.
c
Resistance is shown for 2 standard models. Resistance is proportional to length: long
lengths have higher resistance that may limit the amplitude.
6 English
3487A, 3887, 3888 2004-05
198855002
Rev A
Medtronic Confidential
a
for lead Models: 3487A, 3887, and 3888.
Model 3487A
Model 3887
Quadripolar,
Quadripolar,
in-line
in-line
Straight
Straight
<125 Ω (28 cm)
<125 Ω (28 cm)
<250 Ω (56 cm)
<250 Ω (56 cm)
10 – 110 cm
10 – 110 cm
1.3 mm
1.3 mm
4
4
Cylindrical
Cylindrical
3.0 mm
3.0 mm
6.0 mm
4.0 mm
30.0 mm
24.0 mm
Not applicable
Not applicable
No
No
2.3 mm
2.3 mm
15.2 mm
15.2 mm
40.0 mm
40.0 mm
NeuroLdExt_R01
Model 3888
Quadripolar,
in-line
Straight
<110 Ω (28 cm)
<220 Ω (56 cm)
10 – 110 cm
1.3 mm
4
Cylindrical
6.0 mm
12.0 mm
60.0 mm
7.9 mm
Yes
2.3 mm
15.2 mm
40.0 mm