acumed Polarus Instrucciones De Uso página 50

3 solution plates and nails
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

  • MEXICANO, página 52
EN
Instructions for use
Storage Case Part
Numbers:
Condition
1
:
Exposure Temperature
and Time:
Exposure Temperature
and Time
2
:
Dry Time:
1
Use sterilization packaging/wrap and other accessories appropriate for the cycle parameters recommended in these
instructions and in accordance with national regulations. Refer to PKGI-76 at
Aesculap® rigid sterilization containers.
2
The devices are compatible with exposure for 18 minutes at 134°C.
* Sterilization was validated using a STERIS Amsco 3023 Vacamatic Prevacuum sterilizer and KimGuard
KC600 One-Step wrap.
POST-STERILIZATION INSPECTION
• Do not store or use sterile devices if they are not dry.
- Moisture supports the survival of microorganisms.
- Moisture remaining on wrapped or contained products after sterilization could compromise the
sterile barrier.
- Moisture can corrode metal and dull sharp edges.
• Inspect the sterile barrier for signs of damage. Do not use the product if the sterile barrier has been
compromised.
Prevacuum Steam Sterilizer Parameters
Base: 80-1928, 80-1933
Case Lid: 80-1927, 80-1932
Wrapped
270°F (132°C)
4 minutes
273°F (134°C)
3 minutes
30 minutes
English (non-U.S.) – EN / PAGE 50
www.acumed.net/ifu
for sterilization in

Publicidad

Tabla de contenido
loading

Productos relacionados para acumed Polarus

Tabla de contenido