ART46983-001_rev-A 24-Aug-2022 2:30pm
Help:
View in-app tutorials, access the product labeling, and review the App's legal
information. You can also view the Event Log, which is a list of events recorded
by the App. This may be used by Customer Service to help troubleshoot.
About:
View App software version and other information.
Maintenance and Disposal
Cleaning and Disinfecting the Reader
Cleaning and disinfecting your Reader is important to prevent the spread of
infectious diseases. The Reader has a mean use life of 3 years and has been
validated for 156 cleaning and disinfection cycles (the equivalent of 1 cycle per
week for 3 years).
You should clean and disinfect the Reader once a week. The Reader should also
be cleaned and disinfected prior to being handled by any person providing
testing assistance to the user.
Cleaning is the physical removal of organic soil from the Reader surfaces.
Keeping the Reader clean helps ensure that it is working properly and that no
dirt gets in the device. Cleaning allows for successful, subsequent disinfection.
Disinfection is a process that destroys pathogens, such as viruses and other
microorganisms, on the Reader surfaces. Disinfecting the Reader helps ensure
that no infection is passed on when you or others come in contact with the
Reader.
This device is not intended for use with multiple patients in health care or
assisted-use settings such as hospitals, physician o"ces, or long-term care
facilities because it has not been cleared by FDA for use in these settings,
including for routine assisted testing or as part of glycemic control procedures.
Use of this device on multiple patients may lead to transmission of Human
Immunodeficiency Virus (HIV), Hepatitis C Virus (HCV), Hepatitis B Virus (HBV), or
other bloodborne pathogens.
To clean and disinfect your Reader, you will need Clorox Healthcare Bleach
Germicidal Wipes, EPA Reg. #67619-12.
These disinfectant wipes contain a 0.55% Sodium Hypochlorite (NaOCl) solution