commercial, or hospital environment.
Immunity test: Surge; IEC 61000-4-5
IEC 60601 test level: ± 1 kV di!erential mode; ± 2 kV common mode
Compliance level: : ± 1 kV di!erential mode; ± 2 kV common mode
Electromagnetic environment – guidance: Mains power quality should be that of a typical domestic
commercial, or hospital environment.
Immunity test: Voltage dips, short interruptions and voltage variations on power supply input lines; IEC
61000-4-11
IEC 60601 test level:
0 % UT ; 0.5 cycle At 0⁰, 45⁰, 90⁰, 135⁰, 180⁰, 225⁰, 270⁰ and 315⁰ 0 % UT ; 1 cycle and 70 % UT ; 25/30
cycles Single phase: at 0⁰ UT ; 250/300 cycle
Compliance level:
0 % UT ; 0.5 cycle At 0⁰, 45⁰, 90⁰, 135⁰, 180⁰, 225⁰, 270⁰ and 315⁰ 0 % UT ; 1 cycle and 70 % UT ; 25/30
cycles Single phase: at 0⁰ UT ; 250/300 cycle
Electromagnetic environment – guidance: Mains power quality should be that of a typical domestic,
commercial, or hospital environment. If the user of the System requires continued operation during
power mains interruptions, it is recommended that the System be powered from an uninterruptible
power supply or a battery.
NOTE UT is the a.c. mains voltage prior to application of the test level
Immunity test: Power frequency (50/60 Hz); magnetic field; IEC 61000-4-8
IEC 60601 test level: 30 A/m
Compliance level: 30 A/m
Electromagnetic environment – guidance: Power frequency magnetic fields should be at levels
characteristic of a typical location in a typical domestic, commercial, or hospital environment.
Immunity test: Conducted RF; IEC 61000-4-6
IEC 60601 test level: 6 Vrms; 150 kHz to 80 MHz
Compliance level: 6 Vrms
Immunity test: Radiated RF; IEC 61000-4-3
IEC 60601 test level: 10 V/m; 80 MHz to 2.7 GHz
Compliance level: 10 V/m
Immunity Test: Proximity fields from RF wireless communications equipment; IEC 61000-4-3
IEC 60601 test level: See the table below