TROUBLESHOOTING
If the measured pressure inside the Heater-Cooler 3T tank is not greater than
21 Pa (see Figure 13), adjust the vacuum regulator of the vacuum source until
the Tank Vacuum gauge indicator reaches the acceptable area of operation
(see Figure 13).
If there is a loss of vacuum during the procedure check that the vacuum source
is operational, the canister is properly sealed, that all connections are tight and
tubing is not kinked.
If the suction canister is completely filled with water, exchange the 3T Aerosol
Collection set to avoid loss of vacuum performance caused by the obstruction
of the hydrophobic filter contacting the water. Exposure of the hydrophobic filter
to water will cause the filter to block flow of air and will reduce and stop vacuum
flow to the canister, reducing the disposable's ability to collect aerosol from the
3T water tanks.
If there is a loss of vacuum during the procedure the air flow to the vacuum
source may be obstructed. Check the vacuum source and the overflow safety
trap for possible obstruction.
If no obstruction is found and a loss of vacuum is still observed, open the
vacuum gauge service port located on the back of the machine (see Figure 14).
Opening this service port by removing the luer lock cap will allow vacuum
source air flow to remove any potential obstruction within the pressure
monitoring line internal to the 3T. Re-attach the luer lock cap and confirm the
obstruction has been removed.
Figure 14) Vacuum Gauge Service Port
If loss of vacuum is still observed apply positive air pressure to the service port
using a sterile syringe to deliver approximately 30cc of air into the service port.
Re-attach the luer lock cap and confirm the obstruction has been removed.

PART NUMBERS

3T Aerosol Collection Set
¼" ID vacuum extension line with connector, 366 cm (12 ft)
The 3T Aerosol Collection Set is supplied non-sterile for 3 calendar days of use.
Discard in accordance with hospital procedures after use.
LABELING
Attention, see instruction for use
Follow instructions for use (white symbol on blue
background)
Batch Code
Purchase order number
Only applies in the U.S.A.:
Sale (and prescription) is restricted to physicians
EA
Quantity
Manufacturer
Use By
6
RETURN OF USED PRODUCT
For Customers within the United States
If for any reason this product must be returned to Sorin Group USA, Inc., a returned
goods authorization (RGA) number is required from Sorin Group USA, Inc., prior to
shipping.
If the product has been in contact with blood or body fluids, it must be thoroughly
cleaned and disinfected before packing. It should be shipped in either the original
carton, or an equivalent carton, to prevent damage during shipment; and it should be
properly labeled with an RGA number and an indication of the biohazardous nature
of the contents of the shipment.
Instructions for cleaning and materials, including appropriate shipping containers,
proper labeling, and an RGA number may be obtained from the Sorin Group USA,
Inc. Returned Goods Coordinator, Quality Assurance Department (800-650-2623).
CAUTION: It is the responsibility of the health care institution to adequately
prepare and identify the product for return shipment. Do not return products
that have been exposed to blood borne infectious diseases.
The shipping address for returned goods is:
Sorin Group USA, Inc.
Returned CV Products
14401 West 65th Way
Arvada, CO 80004-3599
For Customers outside the United States
If for any reason this product must be returned, please contact your sales
050900100
representative for specific instructions.
If the product has been in contact with blood or body fluids, it must be thoroughly
cleaned and disinfected before packing. It should be shipped in either the original
050900111
carton, or an equivalent carton, to prevent damage during shipment.
CAUTION: It is the responsibility of the health care institution to adequately
prepare and identify the product for return shipment. Do not return products
that have been exposed to blood borne infectious diseases.
EN - ENGLISH
Dispose the 3T Aerosol Collection Set within 3
calendar days from the installation date written on
the label
Keep away from heat
Keep dry
3T Aerosol Collection Set • Operating instructions
Copyright
2016
©
SORIN GROUP DEUTSCHLAND GmbH Lindberghstrasse 25
D-80939 Munich, Germany
Tel.:
+49/(0)89/32301-0
Fax:
+49/(0)89/32301-555
All rights reserved, especially the right to reproduction and distribution as
well as translation. No part of this document may be reproduced – by
photocopy, microfilm or any other process – nor may any part of it be
stored, edited, duplicated or distributed by electronic means without the
written permission of SORIN GROUP DEUTSCHLAND GmbH.
The name "Stöckert
" is a Registered Trademark.
®
Caution: Federal (U.S.A.) law restricts this device to sale by or on the
order of a physician.
Manufactured for:
SORIN GROUP DEUTSCHLAND GmbH Lindberghstrasse 25
D-80939 Munich, Germany
Tel.:+49/(0)89/32301-0
Fax:+49/(0)89/32301-555
Distributed in the U.S.A. by: Sorin Group USA, Inc.
14401 W. 65th Way, Arvada, CO 80004,
phone:800.221.7943 phone:303.425.5508
fax:303.467.6584

Publicidad

Tabla de contenido
loading

Tabla de contenido