Device menu

10.4.3.4 Analysis inputs pH / ORP (redox)

Analysis inputs: IN7, IN8, IN9, IN10 equipped with pH / redox / NH3 plug-in card
Go to: Device menu è Configuration è Analogue inputs è Analysis input 1 to 4 è
Configuration
Configuration point Selection / option
Name
Electrode type
Filter time constant
Initial
display range
End of display range
Manual
compensation
Compens.- Temp.
Man. compensation
temperature
Alarm 1
Alarm type
Alarm / event list
Alarm text
Threshold
417102269 Rev. 2-08.2020
Up to 15 characters of text
pH standard
redox
0.0 to 25.0 s
-99999 to +99999 (1)
-99999 to +99999 (1)
Yes
No
Alias values/
Alias value 1: Ch.1 reading
Alias value 2: Ch.1 Temp.
AliasValue 3: Ch.2 reading
* Alias value 4: Ch. 2 Temp.
Alias Value 5: Ch.3 reading
* Alias value 6: Ch. 3 Temp.
AliasValue 7: Ch.4 reading
* Alias value 8: Ch. 4 temp.
* = Please select only the
highlighted points
0 to 99999 (²)
Inactive
Min. alarm
Max.- alarm
alarm window
Alarm window invert.
OFF
Alarm
Event
Up to 20 characters of text
-99999 to +99999
156
Explanation
Measuring channel name
Choosing the type of
measurement
Optimise the reading refresh
The greater the filter time
constant, the longer it takes
to show the reading.
Lower threshold of the
display range
Upper threshold of the
display range
Yes = manual temperature
compensation (fixed stored
value)
No = automatic temperature
compensation (reading)
Source of compensation
temperature with automatic
compensation
Value for man.
Compensation temperature
Choice of alarm type
OFF: No entry
Event: At the end of the
calibration interval, the text,
"Alarm Text" is written in the
event list.
Alarm: At the end of the
calibration interval, the text
"Alarm Text" is written in the
display list.
Text which is displayed
when an alarm is in the
display and/or is on the
event list.
Threshold at which an alarm
is triggered (depending on
the type of alarm)
Default
IN7: Channel 1 pH
IN8: Channel 2 pH
IN9: Channel 3 pH
IN10: Channel 4 pH
pH standard
5.0 s
0,0000 (
1
)
1
200,00 (
)
No
IN7: Alias Value 2:
Channel 1
temperature
IN8: Alias Value 4:
Channel 2
temperature IN9: Alias
Value 6:
Channel 3
temperature
IN 10: Alias value 8:
Channel 4
temperature
25,0 (
2
)
Inactive
Alarm
Alarm 1
Temp.input 1
0,00