10.4.3.6 Analysis inputs measurement range 1 Cr/Ci

Analysis inputs: IN7, IN8, IN9, IN10 equipped with Cr or Ci plug-in card
Go to: Device menu è Configuration è Analogue inputs è Analysis input 1 to 4 è
Measurement range 1
Configuration
point
Unit calculation
Initial display range -99999 to 99999 (
End of display
range
Decimal format
Offset
Alarm 1
Alarm type
Alarm / event list
Alarm text
Threshold
Hysteresis
Window width
Alarm delay on
Alarm delay off
alarm 2
Selection / option
Explanation
µS / cm
Displayed unit
mS / cm
2
Lower threshold of the display area,
)
and assign value for 0 (4) mA with an
analogue output.
Upper limit of the display area, the
-99999 to 99999 (
2
)
same assignment of the value for
20 mA the analogue output.
Auto,
Decimal places of the display
XXXXX
XXXX.X
XXX.XX
XX.XXX
X.XXXX
-999 to +999
Correction value that is added to the
reading. This can, for example,
compensate for measurement errors
caused by line resistances.
Inactive
Choice of alarm type
Min. alarm
Max.- alarm
alarm window
Alarm window invert.
OFF
OFF: No entry
Alarm
Event: At the end of the calibration
Event
interval, the text from "Alarm Text" is
written in the event list.
Alarm: After the end of the calibration,
the text "alarm text" is displayed on the
display
Up to 20 characters of
Text which is displayed when an alarm
text
is in the display and/or is on the event
list.
-99999 to +99999
Threshold at which an alarm is
triggered (depending on the type of
alarm)
0 to 99999
Distance to the threshold
Alarm on =
< Thresh. – Hyst. (With minimum
alarm)
> Thresh. + Hyst. (With max. alarm)
Alarm OFF =
> Threshold (with minimum alarm)
< Threshold (with max. alarm)
0 to 99999
Size of the alert window (with alarm
type "alarm window") or of the window
in which no alarm is triggered
( "inverted alarm window" in the alarm
type)
0 to 999 s
Time delay between occurrence of the
alarm condition and the alarm turning
on
0 to 999 s
Time delay between end of the alarm
set to on and the alarm turning off
159
Device menu
Default
Cr: µS / cm Ci:
mS/cm
2
0,0000 (
)
500,00 (
2
)
XXX.XX
0,00
Inactive
Alarm
Alarm 1
Temp.input 1
0,00
0,00
0,00
0 sec
0 sec
417102269 Rev. 2-08.2020