Customer-Specific Linearisation - Ecolab Versatronic Instrucciones De Servicio

Ocultar thumbs Ver también para Versatronic:
Tabla de contenido

Publicidad

Idiomas disponibles
  • MX

Idiomas disponibles

12.8.8

Customer-specific linearisation

Value tables and formulas for customer-specific linearisation must be created using the
Ecolab PC Setup program. When transferring data to the device, the linearisation is
loaded into Ecolab 'Versatronic' together with the Setup. After the linearisation is loaded
into the device, it can be used by analogue inputs in the configuration of the following
types:
n
Universal inputs with a standard signal (except pH / T-compensated measurement of
chlorine) see
n
Temperature inputs for resistance thermometer with customer-specific characteristic
(up to 400 Ω, 4000 Ω or 100 kΩ)
Ä Chapter 10.4.3.1 'Temperature inputs ' on page 151.
see
n
Analysis measuring inputs for electrolytic conductivity
See
Ä Chapter 10.4.3.5 'Analysis inputs configuration Cr / Ci' on page 158.
Go to settings:
Just Setup è customer-specific linearisation è Tabs 1 to 8
Fig. 127: Customer-specific linearisation
Icon button supporting values table
Function
„"Insert row"
adds a new row above the selected line (
"Delete row"
deletes the selected row from the support values table (
"Delete table"
deletes all entries in the supporting values table
"Generate formula"
generates an approximation formula from the values of the supporting
values table
1)
(
Lines are selected by tapping the line numbers.
Ä Chapter 10.4.3.2 'Universal inputs Configuration' on page 153.
1
)
221
PC setup program
Button
1
)
417102269 Rev. 2-08.2020