Problem Bad Check Valve:
Problem Bad Check Valve:
Chromatogram and Data
Chromatogram and Data
0.06
0.04
0.02
0.00
0.020
0.010
0.000
0.06
0.04
0.02
0.00
0.00
0.50
1.00
1.50
•System Benchmarking data
Acetone
Retention
Area
Average
0.323
52245
%RSD
0.69
Naphthalene Retention
Area
Average
1.633
155515
%RSD
0.44
Acenaphthene Retention
Area
Average
2.893
101332
%RSD
0.44
•Good Check Valve
•Bad Check Valve
•Good Check Valve
2.00
2.50
3.00
3.50
Minutes
N=45
Plate
Tailing
Count
1.18
3144
Plate
Tailing
Count
1.13
11009
Plate
Tailing
Count
1.08
10436
7000.00
6500.00
6000.00
7000.00
6500.00
6000.00
7000.00
6500.00
4.00
6000.00
0.50
1.00
•Benchmarking data + 9 injections
Acetone
Average
%RSD
Naphthalene Retention
Average
%RSD
Acenaphthene Retention
Average
%RSD
•~6600 psi
•~6800 psi
•~6600 psi
•~6600 psi
1.50
2.00
2.50
3.00
Minutes
(after fix)N=54
Retention
Area
Tailing
0.323
50804
1.18
0.68
Area
Tailing
1.631
153239
1.12
0.47
Area
Tailing
2.893
99538
1.08
0.44
©2007 Waters Corporation
3.50
4.00
Plate
Count
3152
Plate
Count
10753
Plate
Count
10220