Problem – – Sample Draw Volume
Problem
Reproducibility - -
Reproducibility
Symptom:
Peak area increases after first injection from the same vial (first injection –
low, latter injections OK)
—
Possible cause:
FIRST Time.
"Vacuum formation"
injection needle.
—
Vacuum draws some sample back out of the needle
Situation can be aggravated by over filling the vial
Situation can be aggravated by over filling the vial
o
o
(Never fill the vial all the way to the top)
—
Test:
Remove cap and septum* from the vial.Perform multiple injections. Measure
peak area to determine if the septum/cap is the
•*
Some auto samplers may have to have the vial sensor disabled
•Vial A
•Inj. 1
•Area = 200,000
Sample Draw Volume
Venting
Venting
Inadequate venting upon needle piercing the septum/cap for the
caused by the septum/cap sealing around the
cause.
•Vial A
•Inj.2
•Area = 250,000
©2007 Waters Corporation
114