Product description: VERSAJET II™ console - model 66800039
Classification:
class I / type BF equipment
Requirements:
IEC 60601-1
Table 2
Test
Ground integrity
Earth leakage
Earth leakage
Enclosure leakage
Enclosure leakage
Patient leakage
Patient leakage
Input VAC applied to
patient applied part
Notes:
• For earth leakage current, single-fault condition shall mean the interruption of either power supply conductor,
one at a time.
• For enclosure leakage current or patient leakage current, single-fault condition shall mean the interruption of
either power supply conductor or the protective earth conductor, one at a time.
• For patient leakage current, single-fault condition shall also mean application of rated mains voltage to the
patient applied part relative to the protective earth conductor.
Important note
The VERSAJET II console is NOT user serviceable and under no situation should the casing be opened. Any
attempt to open the unit by the user will void all warranties and render the user responsible for all subsequent
repairs to the unit. Contact your local Smith & Nephew customer care representative or sales representative to
resolve all console problems.
Equipment Condition
Normal
Normal
Single-fault
Normal
Single-fault
Normal
Single-fault
Single-fault
Limit at 120 V
0.2 Ohms
< 250 µ Amp
< 500 µ Amp
< 50 µ Amp
< 250 µ Amp
< 50 µ Amp
< 250 µ Amp
< 2500 µ Amp
26
Limit at 240V
0.2 Ohms
< 500 µ Amp
< 1000 µ Amp
< 100 µ Amp
< 500 µ Amp
< 100 µ Amp
< 500 µ Amp
< 5000 µ Amp