Information for Prescribers
Guidance and manufacturer's declaration – electromagnetic immunity
Voltage dips, short
interruptions and
voltage variations on
power supply input
lines IEC 61000-4-11
(Programming
Wand only)
Power frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
NOTE U
is the a.c. mains voltage prior to application of the test level.
T
Precision Montage™ MRI System Information for Prescribers
91053246-02 12 of 397
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
30 A/m
<5 % U
T
(>95 % dip in U
)
T
for 0,5 cycle
40 % U
T
(60 % dip in U
)
T
for 5 cycles
70 % U
T
(30 % dip in U
)
T
for 25 cycles
<5 % U
T
(>95 % dip in U
)
T
for 5 s
30 A/m
Mains power quality
should be that of a
typical commercial or
hospital environment.
If the user of the
Precision Montage
MRI System requires
continued operation
during power mains
interruptions, it
is recommended
that the Precision
Montage MRI
System be
powered from an
uninterruptible power
supply or a battery.
Power frequency
magnetic fields
should be at levels
characteristic of a
typical location in a
typical commercial or
hospital environment.
Magnetic fields from
common appliances
are not expected to
affect the device.