Caution, hot surface
Protect from moisture
16.2 Operating states/error signals
The prosthesis indicates operating states and error messages through beeps and vibration sig
nals.
16.2.1 Signals for operating states
Battery charger connected/disconnected
Beep signal
1x short
Mode switching
INFORMATION
When mute mode (silent mode) is activated, no beep and vibration signals are generated.
INFORMATION
If the Volume parameter is set to '0' in the Cockpit app, there are no beep signals (see
Page 84).
Beep signal
Vibration
signal
1x short
1x short
1x long
1x long
1x short
1x short
2x short
2x short
3x short
3x short
Vibration signal
3x long
1x before beep signal Self-test completed successfully, product is
Additional action per
formed
Mode switching using the
Cockpit app
Bouncing on the forefoot fol
lowed by unloading the
prosthetic leg
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Weight taken off prosthetic
leg and leg kept still for 1
second
Event
Charging mode started (3 sec. after connecting
the battery charger)
operational
Event
Mode switching is performed
using the Cockpit app.
Bouncing pattern recognised.
Switching to basic mode
(mode 1) carried out.
Switching to MyMode 1 (mode 2)
carried out.
Switching to MyMode 2 (mode 3)
carried out.
99