USER'S MANUAL
ERROR MESSAGES
DISPLAY
DESCRIPTION OF ERROR
Error 1
Overheating / overloading of electronics
Pedal of the foot control pressed during
Error 2
switching on
Error 3
Display keys pressed during switching on
Error 4
Running time limiter as a result of
Error 5
15 minutes of continuous operation
Error with » Electric motor /
Error 6
Reboot
Error
e.g.: 05 6303
If the error messages described cannot be resolved
a check by an authorised service partner is required.
For a list of iM3 Authorised Service Partners in your country
please visit:
Europe & UK: www.im3vet.com
USA: www.im3vet.com
Australia/Asia: www.im3vet.com.au
EVOLUTION VETERINARY DENTAL UNITS
Endo contra-angle handpiece /
Piezo scaler«
SOLUTIONS
> ➊ Disconnect device from power supply
➋ Wait 5 minutes and allow the system to cool
➌ Switch the device back on again and restart
the function
> Do not press the pedal of the foot control
> Do not press the display keys
> Check the pedal of the foot control
> Do not press the pedal of the foot control any more
[release completely]
> ➊ Check the electric motor / Endo contra-angle
handpiece / scaler is correctly attached to the
coupling
> ➋ electric motor / Endo contra-angle handpiece /
replace tip
> Switch the power supply off and on again
> Restart system
> Contact an authorised iM3 service partner
immediately
22 � EN