General information
»
Comply with all national directives, standards and
specifications for the cleaning, disinfection and
sterilisation of medical products as well as the specific
specifications for dental practices and clinics.
»
When selecting the cleaning and disinfecting agents to
be used, observe the specified information (see "Manual
cleaning, intermediate rinsing, disinfection, final rinse,
drying" and "Automatic cleaning, intermediate rinsing,
disinfection, final rinse, drying").
»
Comply with the concentrations, temperatures,
residence times and post-rinsing specifications issued by
the manufacturer of the cleaning and disinfectant agent.
»
Only use cleaning agents that are non-protein-fixing,
aldehyde-free and display material compatibility with
the product.
»
Only use disinfectants that are aldehyde-free and display
material compatibility with the product.
»
Do not use any rinse aid (danger of toxic residue on the
components).
»
Only use freshly-produced solutions.
»
Only use distilled or deionised water (not sterile water)
with a low bacterial count (at least drinking water
quality) that is free from facultative pathogenic micro-
organisms (e.g. legionella bacteria).
»
Use clean, dry, oil and particle-free compressed air.
»
Do not exceed temperatures of 93 °C.
»
All devices used (e.g. ultrasonic bath, cleaning and
disinfection device (washer-disinfector) must be
regularly serviced and inspected.
7.3 Preparation at the operating location
Wear protective gloves
Wear protective goggles.
Use mouth and nose protection
Use protective clothing
EVOLUTION VETERINARY DENTAL UNITS
»
»
»
»
»
Fig.39 - Riser piper
WARNING
Risk of infection from contaminated products.
Risk of cross contamination
» Prepare the product correctly and promptly
before its first use.
Unscrew the bottle from the bottle receptacle [A].
Remove the riser pipe from the bottle receptacle [B].
Thoroughly wipe down the exterior surfaces of all
components with cleaning wipes to remove coarse
organic and inorganic soiling:
»
one cleaning wipe for the small components, e.g.
the individual parts of the riser pipe and
»
two cleaning wipes for larger components, e.g.
the bottle.
Note the action time of the cleaning agent.
Protect the unit from contamination when transporting
it from the treatment chair to the reprocessing location.
40 � EN